Related Experiment Video
Updated: Sep 20, 2026

Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
Factors which may increase stresses at the pin-bone interface in external fixation: a finite element analysis study
1Department of Orthopaedic Surgery, The Glenfield Hospital, Groby Road, Leicester LE3 9QP.
Abstract:
The pin-bone interface of the external fixation system has been studied using a finite element mode. The maximum stresses were observed in the near 'cortex' adjacent to the site of entry of the pin. This location coincides with the area where loosening is usually first observed radiologically. Other results indicate that the stress values are significantly increased by using deep threads and by using stainless steel instead of titanium.
More Related Videos
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
Related Concept Videos
Applications of Stress
The...
Stress Concentrations in Circular Shafts
Components of Stress
Interestingly, the hidden cube faces also experience these stresses, equal and opposite to those on the...
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes.
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress Concentrations