Intravascular ultrasound based dose assessment in endovascular brachytherapy
Gianpiero Catalano1, Vittorio Tamburini, Antonio Colombo
1Division of Radiotherapy, European Institute of Oncology, via Ripamonti 435, 20141 Milan, Italy.
Background:
the role of endovascular brachytherapy in restenosis prevention is well documented. Dose is usually prescribed at a fixed distance from the source axis by angiographic quantification of vessel diameter. Recently, intravascular ultrasound (IVUS) was introduced in dose prescription, allowing a better evaluation of the vessel anatomy. This study retrospectively explores the difference between prescription following angiographic vessel sizing and delivered dose calculated with IVUS.
Methods And Results:
Seventeen lesions were studied with IVUS, identifying on irradiated segment, three sections on which measuring minimal and maximal distance from the centre of IVUS catheter to the adventitia; using dedicated software, corresponding doses were calculated. The dose ranged widely, with maximal and minimal values of 71.6 and 4.9 Gy; furthermore, heterogeneity in dose among different sections was observed. In the central section, the maximal dose was 206% of the one prescribed with the QCA model at 2 mm from the source axis, while the minimal dose was 96%. In proximal and distal sections, respective values were 182, 45, 243, and 122%.
Conclusions:
Our analysis confirmed the dose inhomogeneity delivered with an angiographic fixed-dose prescription strategy. A dose variation was found along the irradiated segment due to the differences in vessel thickness. IVUS emerged as an important tool in endovascular brachytherapy, especially for irregular-shaped vessels.