Fluorometric determination of urea by flow injection analysis
M S Abdel-Latif1, G G Guilbault
1Department of Chemistry, University of New Orleans, Louisiana 70148.
Abstract:
Urea was determined using fluorometry with flow injection analysis. O-phthalaldehyde (OPA) reacts with enzymatically generated ammonia and sulfite in alkaline medium to give a highly fluorescent compound that has an excitation wavelength of 372 nm and an emission wavelength of about 430 nm. The method is more selective to ammonia than the one which uses mercaptoethanol in place of the sulfite. Urease was immobilized to a Pall Immunodyne membrane which is commercially available. The immobilization occurs through covalent bonding which results in a highly stable enzyme preparation. The enzymatic membrane was fitted in a 5 cm long, 0.125 inch o.d. Teflon tubing which served as the enzymatic reactor. The system is difficult to use for the analysis of urea in serum because some compounds normally present in serum fluoresce at the same wavelength. This results in higher values for urea. If the reaction system is to be used for the evaluation of urea in serum, a blank should be run so that urea concentration can be calculated by difference.