Related Experiment Videos
Accuracy of the endpoint assay for virus titration
L K Nielsen1, G K Smyth, P F Greenfield
1Department of Chemical Engineering, University of Queensland, Australia.
Cytotechnology
|January 1, 1992
Abstract:
The statistics of estimators used with the endpoint assay for virus titration were investigated. For a standard assay with 10 wells/dilution, the graphical estimator traditionally used was found to produce estimates with significant positive bias and a relatively low accuracy. Furthermore, the graphical estimator was found to be inconsistent. A superior estimator based on the maximum likelihood principle was developed. The results are discussed in relation to the choice between the endpoint titration assay and the plaque assay, and an alternative two-stage assay is presented.