A new pacemaker algorithm for continuous capture verification and automatic threshold determination: elimination of

G K Feld1, C J Love, J Camerlo

  • 1Department of Medicine, University of California, San Diego.

Pacing and Clinical Electrophysiology : PACE
|February 1, 1992
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Equine allergic skin diseases: Clinical consensus guidelines of the World Association for Veterinary Dermatology.

Veterinary dermatology·2023
Same author

Anisotropic Absorption of Pure Spin Currents.

Physical review letters·2016
Same author

First case report of ultrastructural cutaneous abnormalities in equine atopic dermatitis.

Research in veterinary science·2014
Same author

Pilot study to evaluate the effect of topical dimethicone on clinical signs and skin barrier function in dogs with naturally occurring atopic dermatitis.

Veterinary medicine international·2013
Same author

A young woman with syncope and sudden death What molecular mechanisms could be involved?

Trends in cardiovascular medicine·2011
Same author

A Scoring Index for Clinical Signs of Flea Allergy Dermatitis in the cat.

Veterinary therapeutics : research in applied veterinary medicine·2009

A novel pacemaker algorithm automatically verifies capture and determines pacing thresholds by detecting evoked potentials. This technology ensures reliable pacemaker function by continuously monitoring and adjusting output, enhancing patient safety.

Area of Science:

  • Biomedical Engineering
  • Cardiovascular Technology
  • Implantable Devices

Background:

  • Pacemaker implantation requires reliable verification of capture and accurate determination of pacing thresholds.
  • Current methods for assessing pacemaker capture can be manual and time-consuming.
  • Ensuring consistent pacemaker output is critical for patient safety and effective therapy.

Purpose of the Study:

  • To evaluate a new pacemaker algorithm for automatic verification of pacemaker capture.
  • To assess the algorithm's ability to determine pacing thresholds using stimulus evoked potentials.
  • To investigate the algorithm's performance in both bipolar and unipolar pacing modes.

Main Methods:

  • A triphasic, charge-balanced pacing pulse was generated using a hardware feedback circuit and software template matching.

Related Experiment Videos

  • A capture window was defined by the integral of the evoked depolarization, and capture threshold by one-third of this amplitude.
  • The algorithm measured residual artifact and automatically adjusted pacemaker output with a safety margin.
  • Main Results:

    • The algorithm successfully eliminated pacing stimulus afterpotential and detected evoked responses.
    • Loss of capture was immediately detected in all patients during simulated events.
    • Data showed consistency between algorithm runs, with no significant differences in key parameters between initial runs and post-simulation.

    Conclusions:

    • The developed algorithm reliably and automatically verifies pacemaker capture and determines pacing thresholds.
    • This automated approach enhances patient safety by ensuring consistent and appropriate pacemaker output.
    • The algorithm demonstrates effective performance in both bipolar and unipolar pacing configurations.