A multicenter experience with a bipolar tined polyurethane ventricular lead.

D L Hayes1, K J Graham, M Irwin

  • 1Division of Cardiovascular Diseases and Internal Medicine, Mayo Clinic, Rochester, Minnesota 55905.

Pacing and Clinical Electrophysiology : PACE
|July 1, 1992
PubMed
Summary

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

The Way It Was.

Annual review of microbiology·2019
Same author

DNA Damage Repair in Huntington's Disease and Other Neurodegenerative Diseases.

Neurotherapeutics : the journal of the American Society for Experimental NeuroTherapeutics·2019
Same author

The Surgical Management of Ruptured Abdominal Aortic Aneurysm: A Ten -Year Experience.

The Australian and New Zealand journal of surgery·2017
Same author

Socioeconomic status and lifestyle behaviours in cancer survivors: smoking and physical activity.

Current oncology (Toronto, Ont.)·2017
Same author

A Supernova at z = 0.458 and Cosmologya.

Annals of the New York Academy of Sciences·2015
Same author

Reproductive biology of the greeneye spurdog Squalus chloroculus (Squaliformes, Squalidae).

Journal of fish biology·2015

The Medtronic 4012 pacing lead shows an unacceptable failure rate of 20.9% at 6 years due to insulation issues. Pacemaker-dependent patients may need lead replacement, while others require close monitoring.

Area of Science:

  • Cardiovascular Medicine
  • Biomedical Engineering
  • Medical Device Technology

Background:

  • A multicenter study evaluated the failure rate of the Medtronic 4012 polyurethane bipolar tined pacing lead.
  • 1,190 leads were implanted across six centers in the US and Canada to assess lead failure probability and manifestations.

Discussion:

  • Failures were specifically linked to insulation problems, manifesting as sensing abnormalities, capture failure, and early battery depletion.
  • A significant decrease in measured impedance was also a key indicator of lead malfunction.
  • Of 95 definite failures, 16.8% presented with symptoms mirroring pre-pacemaker placement conditions.

Key Insights:

  • Kaplan-Meier analysis revealed a 20.9% probability of Medtronic 4012 lead failure by 6 years post-implantation.

Related Experiment Videos

  • The observed failure rate is deemed unacceptable by the study authors.
  • Insulation failure is the primary concern for this specific pacing lead model.
  • Outlook:

    • Strong consideration for lead replacement in pacemaker-dependent patients is recommended.
    • Close monitoring of non-dependent patients is advised due to the high failure rate.
    • Further research into lead material durability and design improvements is warranted.