Incidence and timing of activity parameter changes in activity responsive pacing systems.
T Ahern1, C Nydegger, D J McCormick
1Cardiac Electrophysiology Laboratory, Hahnemann University, Philadelphia, Pennsylvania.
Pacing and Clinical Electrophysiology : PACE
|May 1, 1992
Summary
Reprogramming activity parameters in rate-responsive pacemakers (VVIR and DDDR) showed early changes in DDDR systems, suggesting predischarge programming may not predict long-term needs. Early sensor evaluation is recommended.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Device Technology
Background:
- Activity-responsive pacing systems aim to optimize heart rate based on physiological demand.
- Rate response parameter reprogramming is crucial for effective pacemaker function post-implantation.
- Two common systems, VVI,R (Activitrax) and DDD,R (Synchrony), were compared.
Purpose of the Study:
- To evaluate the incidence and timing of rate response parameter reprogramming in VVI,R and DDD,R pacemakers during the first year post-implantation.
- To compare reprogramming strategies based on symptoms versus objective sensor data.
- To assess the predictive value of initial programming for long-term rate response requirements.
Main Methods:
- A comparative study involving 24 patients with VVI,R systems and 21 patients with DDD,R systems.
- Activitrax patients' parameters were adjusted based on symptoms.
- Synchrony patients' parameters were adjusted using sensor data from slow and fast walk protocols.
Main Results:
- No significant difference in reprogramming incidence over the year for Activitrax patients.
- Synchrony patients showed a higher incidence of reprogramming at 1 month compared to later visits (P < 0.02).
- Early post-implant sensor data (4-6 weeks) revealed alterations in sensor response, potentially due to local changes at the pacemaker site.
Conclusions:
- Predischarge programming may not accurately reflect long-term rate response needs in activity-responsive pacing systems.
- Early evaluation of sensor function using an exercise protocol (4-6 weeks post-implant) is recommended for all pacemakers using piezoelectric crystals.
- Objective data and serial evaluation are valuable for optimizing pacemaker settings.