Alterations of phasic coronary artery flow velocity in humans during percutaneous coronary angioplasty

J Segal1, M J Kern, N A Scott

  • 1Division of Cardiology, George Washington University, Washington, D.C.

Insights

Doppler guide wire measurements of coronary artery blood flow velocity and diastolic/systolic velocity ratio distal to stenosis are more relevant than flow reserve for assessing angioplasty outcomes. These distal measurements reflect immediate physiologic changes after coronary angioplasty.

Area of Science:

  • Cardiovascular medicine
  • Interventional cardiology
  • Diagnostic imaging

Background:

  • Previous Doppler catheter studies assessing proximal coronary arteries after angioplasty showed limited success in demonstrating immediate improvement.
  • Measuring blood flow velocity and reserve distal to a stenosis may offer insights into the physiologic significance of coronary artery disease.

Purpose of the Study:

  • To measure blood flow velocity, flow reserve, and diastolic/systolic velocity ratio both proximal and distal to coronary artery stenosis.
  • To evaluate the utility of these Doppler-derived parameters in assessing the immediate outcome of coronary angioplasty.

Main Methods:

  • A low-profile Doppler angioplasty guide wire was used to obtain spectral flow velocity data.
  • Measurements were performed proximal and distal to stenosis in the left anterior descending or left circumflex coronary arteries.
  • The study included 38 patients undergoing angioplasty and 12 controls without significant coronary artery disease.

Main Results:

  • Significant improvement in mean peak velocity was observed in distal coronary arteries post-angioplasty (19 to 35 cm/s).
  • Proximal peak velocity increases were less pronounced.
  • Mean flow reserve remained unchanged both proximally and distally.
  • The distal diastolic/systolic velocity ratio significantly increased post-angioplasty (1.3 to 1.9), returning to normal patterns.

Conclusions:

  • Doppler guide wire flow velocity measurements can be performed distal to coronary stenosis.
  • Distal phasic velocity patterns and diastolic/systolic velocity ratio changes are more indicative of immediate angioplasty success than flow reserve.
  • Proximal and distal flow reserve measurements immediately post-angioplasty had limited utility.
Abstract