Related Experiment Videos
Determination of basal dose rates in Paris system of interstitial implants
Abstract:
The isodose distributions are related to the source configurations in the Paris system of interstitial implants. In this system, the basal dose rates are determined at those points identified relative to the implant. Once these points have been located, the basal dose rates are merely the summation of the dose rate contributions from all the sources. By applying the symmetry properties of the implant, the number of basal dose rate computations and the number of dose rate summations are reduced. The reduction in both the number of computations and the number of summations has facilitated the determination of the average basal dose rate. After the average basal dose rate has been determined, the reference dose rate that is used clinically is calculated. Examples of determining the basal dose rates for a few interstitial implants are presented.