Is blood pressure measurement with a standard cuff reliable?
1Department of Development of Primary Health Care, County Council West, Stockholm, Sweden.
Abstract:
This study compares blood pressures measured with a standard cuff (rubber bag 12 x 35 cm) with concomitantly measured intra-arterial pressures in 48 subjects. With the standard cuff, and using a diagnostic cut-off limit for diastolic hypertension of 90 mmHg, 15/48 patients were found to be hypertensive, whereas only 4/48 had intraarterial pressure above 90 mmHg. Thus, the specificity of the non-invasive method was only 75%. If higher diagnostic cut-off limits were used, e.g. 95 mmHg, specificity increased to 84%, and with lower cut-offs specificity was 63% for 85 mmHg and 52% for 80 mmHg. A specificity of only 75% is very poor for a method commonly used in screening examinations and may lead to considerable over-diagnosis of mild hypertension.