[The tilted screen-cassette ("grid cut-off" effect). A problem of bedside thoracic diagnosis]
U Fischer1, R Vosshenrich, E Grabbe
1Abteilung für Röntgendiagnostik I, Universität Göttingen.
Abstract:
Tilting of a grid during portable radiography leads to uneven exposures, and errors greater than 3 degrees can lead to errors in interpretation. Differentiation from abnormal findings can be made by recognising exposure difference of extrathoracic comparable areas. The difficulties caused by tilting of the grid can be reduced by increasing the film focus distance and by using suitable grids. A new cassette holder with an integrated balance makes it possible to correct tilting of the grid rapidly and effectively. This results in improved image quality which can be applied not only to conventional exposure systems but is also of advantage when using digital methods.