Aqueous humor hydrogen peroxide analysis with dichlorophenol-indophenol
S García-Castiñeiras1, S Velázquez, P Martínez
1Department of Ophthalmology, School of Medicine, University of Puerto Rico, San Juan.
Experimental Eye Research
|July 1, 1992
Summary
Hydrogen peroxide (H2O2) is present in aqueous humor, but the dichlorophenol-indophenol assay overestimates levels due to spontaneous H2O2 generation and ascorbate interference. Careful method adjustments are crucial for accurate H2O2 measurements.
Area of Science:
- Biochemistry
- Analytical Chemistry
- Ophthalmology
Background:
- Hydrogen peroxide (H2O2) has been reported as a normal component of aqueous humor.
- The dichlorophenol-indophenol assay is a common method for H2O2 quantification.
- Previous studies have relied on this method without fully addressing its limitations.
Purpose of the Study:
- To critically evaluate the dichlorophenol-indophenol assay for measuring H2O2 in aqueous humor.
- To identify and quantify sources of error in the assay, specifically spontaneous H2O2 generation and ascorbate interference.
- To determine the accuracy of previously reported H2O2 concentrations in aqueous humor.
Main Methods:
- Analysis of aqueous humor using the dichlorophenol-indophenol method.
- Quantification of spontaneous H2O2 generation by monitoring absorbance changes in the absence of sample.
- Assessment of ascorbate's role in the assay and its interference with H2O2 measurement.
- Evaluation of aqueous humor's effect on the auto-oxidation rate of dichlorophenol-indophenol.
Main Results:
- The dichlorophenol-indophenol method exhibits spontaneous H2O2 generation (0.03 nmol min-1 microM-1) independent of ascorbate.
- Ascorbate auto-oxidation is not the source of spontaneous H2O2; it is consumed by dichlorophenol-indophenol.
- Assay results show an artifactual positive correlation between ascorbate and H2O2 levels.
- Rabbit aqueous humor H2O2 measurements using this method, without corrections, yield inaccurate high values (e.g., 25.3 microM).
Conclusions:
- The dichlorophenol-indophenol assay has significant limitations, including spontaneous H2O2 generation and ascorbate interference, leading to overestimation.
- Previous H2O2 concentrations reported for aqueous humor using this method are likely incorrect.
- Appropriate corrections for spontaneous H2O2 generation and ascorbate levels are necessary for accurate H2O2 quantification in aqueous humor.