Related Experiment Videos
CALIBRATION OF A MODIFIED ANDERSEN BACTERIAL AEROSOL SAMPLER.
Applied Microbiology
|January 1, 1964
Summary
The Andersen sampler had issues collecting larger airborne particles. Redesigning its stages and adding a new one improved particle sampling efficiency and range.
Area of Science:
- Environmental Science
- Aerosol Science
- Microbiology
Background:
- The Andersen sampler is a widely used device for assessing airborne microbial contamination.
- Previous versions exhibited limitations in accurately sampling larger airborne particles, potentially skewing results.
Purpose of the Study:
- To identify and rectify defects in the commercial Andersen sampler's particle sampling capabilities.
- To enhance the instrument's performance for more accurate airborne particle size distribution analysis.
Main Methods:
- Investigated the flow regime within the commercial Andersen sampler.
- Redesigned the hole patterns of the upper stages and incorporated an additional stage.
- Developed a new, rational hole pattern for the lower stages.
- Introduced a colony-counting mask for improved assay efficiency.
- Calibrated the modified Andersen sampler system.
Main Results:
- Identified specific defects in the sampling of larger airborne particles.
- Achieved satisfactory sampling performance through redesigning hole patterns and adding a stage.
- Extended the effective sampling range of the instrument.
- Demonstrated the utility of a colony-counting mask for easier analysis.
- Presented a calibration method enabling particle size distribution curve generation.
Conclusions:
- The modified Andersen sampler design significantly improves the sampling of larger airborne particles.
- The enhanced instrument provides a more accurate assessment of airborne particle size distribution.
- The suggested modifications offer a practical solution for improving aerosol sampling accuracy.