Related Experiment Videos
Analysis of unconjugated bilirubin in serum by reversed-phase high performance liquid chromatography
1Department of Clinical Chemistry, Rigshospitalet, University Hospital, Copenhagen, Denmark.
Abstract:
A high performance liquid chromatographic method for the analysis of unconjugated bilirubin in neonatal serum is presented. Bilirubin was dissociated from protein with caffeine reagent and extracted with chloroform. An isocratic, reversed-phase HPLC system based on a C18 column was used. Bilirubin was detected at 450 nm. Bilirubin SRM 916a from National Institute of Standards and Technology, USA was the primary calibrator. An average recovery factor of 0.996 (SD = 0.018; N = 16) was obtained for bilirubin added to neonatal cord sera. The measurement range extended from 25 to 500 mumol l-1 L-1. The method is proposed as a reference method for unconjugated bilirubin in neonatal sera.