Low-temperature storage of bone marrow in nitrogen vapor-phase refrigerators: decreased temperature gradients with an
1Clinical Cryobiology Laboratory, Fred Hutchinson Cancer Research Center, Seattle, Washington.
Abstract:
Large temperature gradients may exist in nitrogen vapor-phase refrigerators. Cryopreserved cells stored at higher levels may be considerably warmer than those stored closer to the liquid nitrogen reservoir. To decrease this temperature gradient, racking systems constructed of aluminum were placed in marrow storage refrigerators. The higher heat conductivity of aluminum resulted in a vapor-phase temperature gradient of only 5.9 degrees C at 22.5 inches above the liquid, as compared to the gradient of 86 degrees C seen with steel frames in a similar refrigerator. Temperature fluxes were minimal with lid opening or nitrogen filling. The thicker frame size and loss of the lowest storage level resulted in a storage capacity 63 percent of that achievable with steel frames and liquid immersion. Consumption of nitrogen was estimated to be 174 to 220 percent of the static usage in this model of refrigerator with 6 inches of nitrogen, but comparable to the consumption expected with full immersion of the racking system, regardless of frame construction. These data demonstrate the feasibility of achieving very low, stable, cryogenic temperatures in a vapor-phase refrigerator.
More Related Videos
06:42A Multi-hole Cryovial Eliminates Freezing Artifacts when Muscle Tissues are Directly Immersed in Liquid Nitrogen
Published on: April 6, 2017
06:42Cryogenic Sample Loading into a Magic Angle Spinning Nuclear Magnetic Resonance Spectrometer that Preserves Cellular Viability
Published on: September 1, 2020