Electrode causation of pacemaker inhibition.

S Furman, V De Caprio

    Chest
    |July 1, 1977
    PubMed
    Summary

    Movement of intracardiac metal portions of an electrode can cause false pacemaker recycling. These electromechanical signals mimic the natural heart rhythm and stop when the electrode movement ceases.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Multidimensional gain control in image representation and processing in vision.

    Biological cybernetics·2014
    Same author

    Controlled synthesis of a large fraction of metallic single-walled carbon nanotube and semiconducting carbon nanowire networks.

    Nanoscale·2011
    Same author

    Management of potential conflict of interest during publication and presentation.

    Journal of cardiac failure·2002
    Same author

    Recognition and correction of subcuticular malposition of pacemaker pulse generators.

    Pacing and clinical electrophysiology : PACE·2001
    Same author

    The mysterious fate of Hyman's pacemaker.

    Pacing and clinical electrophysiology : PACE·2001
    Same author

    Vasoactive intestinal peptide and related molecules induce nitrite accumulation in the extracellular milieu of rat cerebral cortical cultures.

    Neuroscience letters·2001

    Area of Science:

    • Cardiovascular medicine
    • Biomedical engineering
    • Electrophysiology

    Background:

    • Pacemakers are crucial for managing cardiac arrhythmias.
    • Endocardial electrodes are used for pacemaker sensing and pacing.
    • Nonphysiologic signals can interfere with pacemaker function.

    Observation:

    • Electromechanical signals were generated by the movement of intracardiac metal components of an endocardial grasping electrode.
    • These signals mimicked the amplitude and slew rate of the native ventricular electrogram.
    • The generated signals led to inappropriate pacemaker recycling.

    Findings:

    • The movement of metallic surfaces within the heart created false signals.
    • These signals were indistinguishable from genuine cardiac signals to the pacemaker.
    • Cessation of electrode movement resolved the false recycling events.

    Implications:

    • This highlights a potential source of pacemaker malfunction.
    • Awareness of electromechanical interference is crucial for device management.
    • Further research into signal filtering and lead design may prevent such events.