Do fetal pulse oximetry readings at delivery correlate with cord blood oxygenation and acidaemia?
H McNamara1, D C Chung, R Lilford
1Department of Obstetrics and Gynaecology, University of Leeds, Yorks, UK.
Objective:
To assess the accuracy of a pulse oximeter in the fetus.
Design:
A prospective descriptive study, comparing oxygenation and acidaemia of cord blood with oxygen saturation in the arteries of the fetal scalp measured by a pulse oximeter just before delivery. The data were collected using an experimental pulse oximeter and a sensor specifically adapted to cope with the problems of fetal pulse oximetry.
Setting:
The labour ward in a teaching hospital.
Subjects:
Thirty seven Caucasian fetuses presenting by the vertex in normal uncomplicated labour.
Main Outcome Measure:
Fetal pulse oximetry reading shortly before birth, umbilical cord oxygenation and pH, and Apgar scores.
Results:
Data of sufficient quality were obtained from 28 fetuses. There was a highly significant correlation between pulse oximetry reading and umbilical vein oxygen saturation r = 0.59, P = < 0.001). There was also a significant correlation between the pulse oximeter reading and cord blood pH (vein: r = 0.57, P = 0.002, artery: r = 0.63, P = 0.001). Apgar scores were not related to the oximetry results.
Conclusion:
Pulse oximetry readings reflect fetal oxygenation at birth.
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring pulse
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Fetal Circulation
Two umbilical arteries transport blood from the fetus to the placenta. At the placenta, the blood absorbs oxygen and nutrients while simultaneously eliminating waste products. This oxygen-enriched and nutrient-rich blood then returns to the fetus through one...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...