The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Late disassembly of a modular acetabular component. A case report
1Division of Orthopedic Surgery, Dalhousie University, Halifax, Nova Scotia, Canada.
Late disassembly of metal-backed acetabular components in total hip arthroplasty can occur without dislocation. This complication should be considered in patients experiencing late hip pain.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Medical Device Engineering
Background:
- Metal-backed acetabular components are widely used in total hip arthroplasty.
- Disassembly, a known complication, typically presents with femoral component dislocation or radiographic signs.
- Both preassembled and modular designs are susceptible to this failure mode.
Observation:
- A rare case of late acetabular component disassembly is presented.
- This specific instance occurred without femoral component dislocation.
- No radiographic evidence indicated dissociation prior to the event.
Findings:
- The study highlights a previously underreported presentation of acetabular component disassembly.
- Late-onset pain following total hip arthroplasty may be indicative of this complication.
- This emphasizes the need for thorough differential diagnosis in such cases.
Implications:
- Clinicians should consider acetabular component disassembly in the differential diagnosis of late-onset hip pain.
- This complication warrants consideration alongside infection and aseptic loosening.
- Further research may be needed to understand the mechanisms and prevalence of silent disassembly.