Cardiac cycle phase uncertainty: another source of error in indirect blood pressure measurement
1Armstrong Laboratory, Brooks AFB, TX 78235.
Abstract:
Cardiac cycle phase uncertainty causes a small error in indirect estimation of arterial blood pressure by sphygmomanometry. A simple analysis yields a statistical description of the error and a rule for its reduction: correct bias by adding to the systolic estimate and subtracting from the diastolic estimate one-half of the per-cardiac-cycle cuff deflation decrement. If several measurements are taken, the least-square estimate is approximately the arithmetic mean of the greatest and least of the bias corrected estimates.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.