Errors and artifacts of Doppler flowmeters and their solution.

K W Johnston, B C Maruzzo, R S Cobbold

    Archives of Surgery (Chicago, Ill. : 1960)
    |November 1, 1977
    PubMed
    Summary

    A new multifilter system improves continuous-wave Doppler ultrasound accuracy for vascular surgeons. This system offers a simple, inexpensive alternative for real-time display of Doppler waveforms and accurate assessment of arterial disease severity.

    Related Concept Videos

    You might also read

    Related Articles

    Articles linked to this work by shared authors, journal, and citation graph.

    Sort by
    Same author

    Carotid artery stenting: to be or not to be?

    The surgeon : journal of the Royal Colleges of Surgeons of Edinburgh and Ireland·2008
    Same author

    A new time-domain narrowband velocity estimation technique for Doppler ultrasound flow imaging. II. Comparative performance assessment.

    IEEE transactions on ultrasonics, ferroelectrics, and frequency control·2008
    Same author

    Intraventricular administration of rt-PA in patients with intraventricular hemorrhage.

    Southern medical journal·2005
    Same author

    C1-C2 transarticular screw fixation for atlantoaxial instability.

    Southern medical journal·2004
    Same author

    Interpretation of three-dimensional structure from two-dimensional endovascular images: implications for educators in vascular surgery.

    Journal of vascular surgery·2004
    Same author

    Vascular surgery training in general surgery residency programs: the Canadian experience.

    Journal of vascular surgery·2003

    Area of Science:

    • Vascular Surgery
    • Medical Ultrasound
    • Biomedical Engineering

    Background:

    • Continuous-wave Doppler ultrasound flowmeters are crucial for vascular surgeons.
    • Current methods using zero-crossing detectors produce erroneous flow velocity waveforms with artifacts.
    • Accurate quantitative flow velocity measurement is essential for diagnosing arterial disease.

    Purpose of the Study:

    • To introduce a multifilter system as an alternative to zero-crossing detectors for continuous-wave Doppler ultrasound.
    • To evaluate the accuracy and real-time display capabilities of the multifilter system.
    • To assess the correlation between pulsatility index derived from multifilter Doppler waveforms and atherosclerosis severity.

    Main Methods:

    • A multifilter system was developed as a replacement for zero-crossing detectors.

    Related Experiment Videos

  • The system allows for real-time display of Doppler waveforms.
  • Pulsatility index was calculated from Doppler waveforms of peripheral arteries in 155 limbs.
  • Main Results:

    • The multifilter system provides a simple, inexpensive alternative to zero-crossing detectors.
    • It enables real-time display of Doppler waveforms, reducing artifacts.
    • Calculated pulsatility index showed an excellent correlation with atherosclerosis severity determined by arteriography.

    Conclusions:

    • The multifilter system significantly improves the accuracy of continuous-wave Doppler ultrasound for quantitative flow velocity measurement.
    • This technology offers a reliable method for real-time assessment of arterial disease severity.
    • The pulsatility index derived from this system is a valuable non-invasive indicator of atherosclerosis.