Related Experiment Video
Updated: Aug 30, 2026

Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018
Fractures at the tip of long-stem prostheses used for revision hip arthroplasty
Paul Zalzal1, Rajiv Gandhi, Danielle Petruccelli
1Department of Orthopaedic Surgery, Mount Sinai Hospital, Toronto, Ontario, Canada.
Abstract:
We report a retrospective analysis of the incidence of intraoperative femur fractures at the tip of the prosthesis in 45 consecutive, uncemented, long (205-mm), straight hydroxyapatite, fully coated stems used for revision arthroplasty of the hip. Five patients were lost to follow-up. Intraoperative fractures of the distal femur occurred in 7 cases (17.5%). In the remaining 33 cases in which no fracture occurred, the tip of the implanted stem was found to be potentially at risk for causing a fracture in 18 (55%) cases. We concluded that uncemented, 205-mm, straight femoral stems should be used with caution in the setting of revision hip surgery. A bowed stem of that length may be a safer alternative. Otherwise, a shorter, straight stem may reduce the risk of complications at the stem tip.