Effects of sterilization on the Tekscan digital pressure sensor
Howard J Agins1, Valerie S Harder, Eugene P Lautenschlager
1Illinois Bone and Joint Institute, 2401 Ravine Way, Glenview, IL 60025, USA.
Abstract:
Investigations into the effects of sterilization on a new biomechanical pressure sensor are necessary before contemplating in vivo use. Ten, designated Experimental, "K-Scan" digital pressure sensor arrays were sterilized with ethylene oxide gas (EtO), and their ability to accurately and reproducibly measure an applied load of 2225 N (500 lb) was assessed. Simultaneously, 10 un-sterilized sensor arrays, designated Control, were assessed. Each array was loaded 10 times inside a two-dimensional curved surface, and all arrays exhibited high reproducibility (coefficients of variation<2.0%). Following sterilization, the Experimental sensors showed a 22.2% average decrease in recorded force, a statistically significant difference from the pre-sterile data (p<0.002). However, when the Experimental sensors were re-calibrated post-sterilization, they showed only a 0.1% average decrease in recorded force, not a statistically significant difference (p>0.05, beta<0.05). Following 1-week storage, trial 2 data of the Control sensors showed a less dramatic yet significant 3.4% average decrease in recorded force when compared to trial 1 data (p<0.02). Control trial 2, once re-calibrated, showed a 0.5% average decrease in recorded force, not a statistically significant difference (p>0.05, beta<0.05). Results suggest that, following EtO sterilization, accurate and reproducible pressure measurements can be obtained from K-Scan sensors when calibration is performed at time of use.
More Related Videos
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Bioreactor Controls-I
Pressure Gauges
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...