Accuracy of analog telephonic stethoscopy for pediatric telecardiology
John M Belmont1, Leone F Mattioli
1Department of Pediatrics, University of Kansas Medical Center, Kansas City, Kansas 66160-7330, USA.
Objective:
Wide-bandwidth electronic stethoscopy is reliable and accurate for pediatric telecardiology. We tested a much less expensive and more convenient system for the same purpose, a narrow-bandwidth telephonic stethoscope (TS).
Methods:
Seventy-six consecutive patients (mean age: 10.0; standard deviation: 6.5 years) in a pediatric cardiology outpatient clinic were studied. One pediatric cardiologist examined the patients with his acoustic stethoscope (AS); a second examined them within a few minutes using a remote TS. A nurse placed the TS chest piece as directed by the remote examiner via intercom, but neither video examination nor conversation with the parent/patient were permitted. Examiners independently recorded the stethoscope findings for all heart sounds, all murmurs, and heart disease (present/absent). TS accuracy was indexed using the kappa statistic for TS/AS agreement and for TS agreement with auscultatory findings predicted from echocardiographic (echo) studies (N = 49).
Results:
TS/AS agreement was satisfactory for presence/absence of heart disease (kappa = 0.63) and for organic, functional, vibratory, diastolic aortic, and diastolic pulmonic murmurs (kappa range: 0.65-0.75). For other specific murmurs and all heart sounds, TS/AS agreement was either unsatisfactory (kappa < or = 0.60) or indeterminate because prevalence was 0. TS-AS agreement improved when the TS was used by the more-experienced TS examiner and with patients at least 5 years of age. When the older children were examined by the more TS-experienced examiner, the TS-echo comparison yielded kappa = 0.90, raw agreement = 0.96, sensitivity = 0.94, and specificity = 1.00.
Conclusions:
In pediatric patients, a narrow-bandwidth telephonic stethoscope can accurately distinguish between functional and organic murmurs and thus can detect heart disease. Accuracy is greatest when the instrument is used by an experienced examiner with patients at least 5 years of age.
Related Concept Videos
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessment of the Cardiovascular System IV: Auscultation
Normal Heart Sounds
S1 (First Heart Sound)-
S1 is made by the closure of the mitral and tricuspid valves (atrioventricular valves), marking the beginning of systole.
S2 (Second Heart Sound)-
S2 is made by the closure of the aortic and pulmonic valves (semilunar valves), marking the end of the systole.
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...
Types of Reports III: Telephone and Verbal Reports
Here's an overview of each type:
Telephone Orders