Related Experiment Video
Updated: Aug 30, 2026

Hickman Catheter Use for Long-Term Vascular Access in a Preclinical Swine Model
Published on: March 31, 2023
Development of a long-lasting ventricular catheter impregnated with a combination of antibiotics
Wolfgang Kohnen1, Christian Kolbenschlag, Susanne Teske-Keiser
1Department of Hygiene and Environmental Medicine, Johannes Gutenberg-University of Mainz, Hochhaus am Augustusplatz, D-55131 Mainz, Germany. kohnen@mail.uni-mainz.de
Abstract:
A ventricular silicone catheter impregnated with a combination of rifampin and a quinolone was developed in order to prevent ventricular shunt related infections. As model substance for the quinolones we used sparfloxacin, because of its specific physicochemical properties resulting in a quantitative detection also in the presence of a second antibiotic. In our study we focused especially on an optimization of the antibiotic release out of the impregnated catheters in order to develop long lasting devices with a broad antimicrobial spectrum. A release-optimized catheter was tested with an in vitro colonization test and additionally with a method developed to examine the spread of bacteria on a catheter surface. In vitro experiments showed that the impregnated catheters reduce the colonization with Staphylococcus epidermidis for at least 1 year and prevent the spread of bacteria along the catheter surface.