Related Experiment Video
Updated: Aug 30, 2026

Interventional Diagnostic Procedure: A Practical Guide for the Assessment of Coronary Vascular Function
Published on: March 15, 2022
Belgian experience with FilterWire EX in the prevention of embolic events during carotid stenting
Marc Bosiers1, Patrick Peeters, Jurgen Verbist
1Department of Vascular Surgery, AZ St-Blasius, Dendermonde. marc.bosiers@pandora.be
Insights
The FilterWire EX effectively captures debris during carotid stenting, preventing embolic events and preserving neurological status in patients undergoing endovascular repair for carotid artery stenosis.
Area of Science:
- Interventional Neurology
- Vascular Surgery
- Medical Device Technology
Background:
- Carotid artery stenosis poses a significant risk of stroke.
- Endovascular interventions, such as stenting, are increasingly used for treatment.
- Cerebral embolic events are a major concern during these procedures.
Purpose of the Study:
- To evaluate the FilterWire EX system's performance in capturing debris.
- To assess its efficacy in preventing embolic events during carotid interventions.
- To determine its role in maintaining neurological status post-procedure.
Main Methods:
- 100 patients with carotid stenosis underwent stenting with FilterWire EX cerebral protection.
- Filter performance was assessed by debris capture and neurological status.
- Technical success and procedural outcomes were recorded.
Main Results:
- FilterWire EX placement and retrieval were successful in 93% of cases.
- Debris was captured in 56.9% of protected procedures.
- A single transient ischemic attack occurred post-procedure; overall procedural success was 98%.
Conclusions:
- The FilterWire EX is effective in preventing embolic events.
- It helps preserve neurological status during carotid artery stenting.
- The device demonstrates utility in endovascular treatment of carotid stenosis.
Purpose:
To evaluate the performance of the FilterWire EX for the capture and removal of debris and prevention of embolic events during endovascular carotid interventions.
Methods:
In the period from October 2001 to May 2002, 100 patients (60 men; mean age 72 years, range 48-90), 69 symptomatic with >or=50% internal carotid artery (ICA) stenosis and 31 asymptomatic with >or=70% stenosis were scheduled for carotid stenting with cerebral protection provided by the FilterWire EX system. Satisfactory performance of the filter was defined as effectiveness in preventing intraprocedural embolic events shown by visible debris in the filter and unchanged neurological status assessed by an independent neurologist.
Results:
The placement and retrieval of the FilterWire EX was technically successful in 93 of 100 patients. Five delivery failures were due to severe ICA angulation; one stenosis was underestimated on preprocedural imaging, and one case was aborted following a major stroke during sheath manipulation before filter introduction. In 56.9% of the protected cases, the filter contained debris. One neurological event occurred 6 hours after the procedure: a transient ischemic attack from which the patient recovered fully after 2 hours. No other procedure-related events were recorded within 30 days of the procedure. The procedural success among the 93 patients stented under FilterWire EX protection was 98% (92% by intention to treat).
Conclusion:
The FilterWire EX appears to be an effective tool in the prevention of embolic events and in preserving neurological status during endovascular treatment of carotid artery stenosis.