Belgian experience with FilterWire EX in the prevention of embolic events during carotid stenting

Marc Bosiers1, Patrick Peeters, Jurgen Verbist

  • 1Department of Vascular Surgery, AZ St-Blasius, Dendermonde. marc.bosiers@pandora.be

Insights

The FilterWire EX effectively captures debris during carotid stenting, preventing embolic events and preserving neurological status in patients undergoing endovascular repair for carotid artery stenosis.

Area of Science:

  • Interventional Neurology
  • Vascular Surgery
  • Medical Device Technology

Background:

  • Carotid artery stenosis poses a significant risk of stroke.
  • Endovascular interventions, such as stenting, are increasingly used for treatment.
  • Cerebral embolic events are a major concern during these procedures.

Purpose of the Study:

  • To evaluate the FilterWire EX system's performance in capturing debris.
  • To assess its efficacy in preventing embolic events during carotid interventions.
  • To determine its role in maintaining neurological status post-procedure.

Main Methods:

  • 100 patients with carotid stenosis underwent stenting with FilterWire EX cerebral protection.
  • Filter performance was assessed by debris capture and neurological status.
  • Technical success and procedural outcomes were recorded.

Main Results:

  • FilterWire EX placement and retrieval were successful in 93% of cases.
  • Debris was captured in 56.9% of protected procedures.
  • A single transient ischemic attack occurred post-procedure; overall procedural success was 98%.

Conclusions:

  • The FilterWire EX is effective in preventing embolic events.
  • It helps preserve neurological status during carotid artery stenting.
  • The device demonstrates utility in endovascular treatment of carotid stenosis.