Related Experiment Video
Updated: Aug 30, 2026

Radiotracer Administration for High Temporal Resolution Positron Emission Tomography of the Human Brain: Application to FDG-fPET
Published on: October 22, 2019
Standardized Uptake Values in 2-Deoxy-2-[18F]Fluoro-D-Glucose with Positron Emission Tomography. Clinical
Bennett B Chin1, Sujatha Lyengar, Beulah Perdue Sabundayo
1Johns Hopkins School of Medicine, Department of Radiology, Division of Nuclear Medicine, Baltimore, MD 21287, USA. bchin@jhmi.edu
Purpose:
To evaluate the clinical significance of differences in 2-deoxy-2-[18F]fluoro-D-glucose positron emission tomography (FDG-PET) lymph node standardized uptake values (SUV) in human immunodeficiency virus (HIV) infection using iterative reconstruction with segmented attenuation correction (IR SAC) compared to filtered back-projection with measured attenuation correction (FBP MAC).
Procedures:
Seven patients with HIV infection and multiple focal lymph node abnormalities were investigated with whole-body FDG-PET. Mean and maximum SUVs from lymph node regions of interest (n = 961) were compared for quantitative differences between reconstruction techniques.
Results:
IR MAC resulted in significantly lower mean SUV [0.06; 95% (confidence interval (CI)) = 0.04-0.07] and maximum SUV (0.82; 95% CI = 0.77-0.88) values compared to FBP MAC. With IR, segmentation of attenuation correction (AC) resulted in significantly higher mean SUV (0.12; 95% CI = 0.11-0.13) and maximum SUV (0.21; 95% CI = 0.18-0.23) values compared to IR MAC. The overall effect of both IR and SAC was a slight but significant increase in mean SUV (0.06; 95% CI = 0.06-0.08; bias = 2.1%) and a significant decrease in maximum SUV (0.62; 95% CI = 0.56-0.67) compared to FBP MAC.
Conclusions:
With our reconstruction parameters, significant differences in mean and maximum SUV values were observed. The magnitude of the mean SUV difference, however, was small. IR SAC is a promising method to accurately quantify standardized uptake values for clinical use.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
07:28Studying Metabolic Brain Connectivity Using 2-Deoxy-2-[18F]Fluoro-D-Glucose Dynamic Positron Emission Tomography at the Single-subject Level
Published on: January 24, 2025
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET