Related Experiment Video
Updated: Aug 30, 2026

In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
In vivo intravascular MR imaging: transvenous technique for arterial wall imaging
Lawrence V Hofmann1, Robert P Liddell, Aravind Arepally
1Russell H. Morgan Department of Radiology and Radiological Science, The Johns Hopkins Medical Institutions, Blalock 545, 600 North Wolfe Street, Baltimore, Maryland 21287, USA. lhofmann@jhmi.edu
Purpose:
To determine, in vivo, the potential for transvenous magnetic resonance (MR) imaging of the arterial wall and to assess appropriate MR pulse sequences for this method.
Materials And Methods:
MR imaging was performed on 19 vessels (right renal artery, N = 9; left renal artery N = 2; external iliac artery, N = 4; abdominal aorta, N = 4) in nine swine. The animals were either low-density lipoprotein receptor knockout (N = 5) or Yucatan mini-pigs fed an atherogenic diet for 6 to 11 weeks (N = 4). The intravascular MR coil/guide wire (IVMRG) (Surgi-Vision, Gaithersburg, MD) was introduced via the external iliac vein into the inferior vena cava (IVC). The following electrocardiograph-gated MR pulse sequences were obtained: T1-weighted precontrast with and without fat saturation and T1-weighted postcontrast with fat saturation. Two observers scored wall signal and conspicuity and classified the vessel as normal, abnormal, or stented. Images were compared with histopathologic findings.
Results:
The T1-weighted precontrast without fat saturation, T1-weighted precontrast with fat saturation, and T1-weighted postcontrast images correlated with histopathologic findings in 12 of 15 vessels, eight of 10 vessels, and 14 of 16 vessels, respectively. Abnormal histopathologic findings included: arterial wall thickening (N = 3), arterial dissection (N = 2), focal fibrous plaque (N = 2), adherent thrombus (N = 1). The T1-weighted postcontrast images were not compromised by artifacts and had the highest score for vessel wall signal and conspicuity. T1-weighted precontrast images were compromised by chemical shift artifact and poor blood suppression. Negligible artifacts were created by the platinum stent.
Conclusion:
The T1-weighted fat saturated postcontrast pulse sequence was superior to other sequences for transvenous MR imaging of the arterial wall.
More Related Videos
07:50Multiphoton Intravital Imaging for Monitoring Leukocyte Recruitment during Arteriogenesis in a Murine Hindlimb Model
Published on: September 30, 2021
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Related Concept Videos
Venous Thrombosis II: Clinical Manifestations and Diagnostic Studies
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI