Background:
Coronary artery bypass surgery is increasingly performed without an extra-corporeal circuit. A heart-lung machine (HLM) is kept on standby for safety reasons, but rarely used. The tubes of the machines are currently discarded after every operation. Costs and waste would be avoided if HLMs could stay on standby for longer periods of time. We therefore investigated the sterility of intra-tube fluid over time.
Materials And Methods:
Four machines were tested. The tube system was fixed and filled with priming solution. Samples of intra-tube fluid and fluid from the fluid bags were taken after 0, 12, 24, 36, 48 and 72 h. Microbiological cultures were performed by direct inoculation and membrane filtration and incubated up to 14 days. Endotoxin levels were also determined.
Results:
Bacteria were grown from two samples only and were considered as contaminants. Endotoxin concentrations never exceeded acceptable levels.
Conclusion:
To keep HLMs on standby for 72 h is probably safe. The prolonged use of HLMs will contribute to cost and waste reduction in open heart surgery.
Related Concept Videos
Holter Monitor: 24-Hour Monitoring
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Pneumothorax-I
Pneumothorax can be even further classified as spontaneous, traumatic, and tension pneumothorax.
Cardiopulmonary Resuscitation II: ACLS Airway Management
Mechanical Ventilation I: Indication and Settings
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: