S Ishikawa1, S Ohmi, K Nakazawa
1Department of Anesthesiology and Critical Care Medicine, Tokyo Medical and Dental University, School of Medicine, 1-5-45 Yushima, Bunkyo-ku, Tokyo 113-8519, Japan.
Purpose:
This study was undertaken to assess the clinical usefulness of a continuous intra-arterial blood gas (CIABG) monitoring system, Paratrend 7, during thoracic surgery.
Methods:
A sensor of the CIABG monitoring system was inserted into the radial artery in 50 patients. During one-lung/differential lung ventilation, arterial blood samples for estimation of blood gases (ABG) were taken every 45-90 min or when clinically needed. The PCO(2), PO(2), and pH values displayed by the CIABG monitor were recorded prior to arterial blood sampling and were compared with the results of ABG analysis. The mean (bias) and the standard deviation (precision) of the differences were calculated from the data for each parameter.
Results:
Ninety-four blood samples were obtained. The correlation between CIABG and ABG measurements was strong for each parameter: r = 0.83 (PCO(2)), 0.89 (PO(2)), 0.74 (pH). The bias +/- precision between the two methods was 0.4 +/- 3.0 mmHg for PCO(2), -6 +/- 47 mmHg for PO(2), -1.2 +/- 27.4% for PO(2) and -0.01 +/- 0.04 for pH. For PO(2) values <150 mmHg, the bias +/- precision was 1 +/- 28 mmHg.
Conclusion:
The agreement between CIABG and ABG measurements was better for PCO(2) and pH than for PO(2). The PO(2) value displayed on the CIABG monitor may not always be reliable during thoracic surgical procedures.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Endoscopic Studies II: Thoracocentesis
Description
Excess pleural fluid or air may accumulate in some respiratory disorders in the thoracic cavity. To treat pleural effusion, a physician conducts thoracentesis by carefully piercing the chest wall and entering...
Blood Studies I: ABG and VBG
Arterial Blood Gas (ABG)
Arterial Blood Gas (ABG) studies are crucial for assessing the lungs' ability to supply oxygen and remove carbon dioxide, reflecting the patient's ventilation status. They also help understand the kidneys' capacity to reabsorb or...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Aneurysm IV: Nursing Management