Slow equilibration of reversed-phase columns for the separation of ionized solutes
D H Marchand1, L A Williams, J W Dolan
1Department of Chemistry, University of Wisconsin River Falls, River Falls, WI 54022, USA. daniel.marchand@uwrf.edu
Abstract:
Reversed-phase columns that have been stored in buffer-free solvents can exhibit pronounced retention-time drift when buffered, low-pH mobile phases are used with ionized solutes. Whereas non-ionized compounds exhibit constant retention times within 20 min of the beginning of mobile phase flow, the retention of ionized compounds can continue to change (by 20% or more) for several hours. If mobile phase pH is changed from low to high and back again, an even longer time may be required before the column reaches equilibration at low pH. The speed of column equilibration for ionized solutes can vary significantly among different reversed-phase columns and is not affected by flow rate.
Related Concept Videos
Ion Exchange
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
Ion-Exchange Chromatography