Tensile strength of 19- and 20-gauge arrow epidural catheters
Ban C H Tsui1, Brendan Finucane
1From the Department of Anesthesiology and Pain Medicine, University of Alberta, Edmonton, Alberta, Canada.
Unlabelled:
There are no data about the mechanical properties of the 20-gauge reinforced Arrow epidural catheter, which has a similar design to the 19-gauge reinforced Arrow catheter. In this study, we compared the mechanical properties of 19- and 20-gauge Arrow epidural catheters at 22 degrees C and 37 degrees C. The distal 12 cm of each catheter was suspended in an enclosed chamber maintained at either 22 degrees C or 37 degrees C. A pair of forceps was applied to each catheter 5 cm from the distal end (fixed site). Another pair of forceps was applied 15 cm from the catheter tip (traction site). The catheter was pulled. At 22 degrees C, the mean fracture force was 2.24 kg (range, 1.96 - 2.41 kg) for 5 19-gauge catheters and 2.17 kg (range, 2.04-2.41 kg) for 5 20-gauge catheters. At 37 degrees C, the mean force was 1.98 kg (range 1.84- 2.15 kg) for 5 19-gauge catheters and 1.99 kg (range, 1.81-2.09 kg) for 5 20-gauge catheters. There were no significant differences in tensile strength between the two different gauge catheters at either temperature. All the 19-gauge catheters fractured at the same "fixed site" at both temperatures. All the 20-gauge catheters elongated at the "fixed site" but fractured at the "traction site." We conclude that using the smaller-gauge catheter (20-gauge) is not associated with a more frequent rate of fracture.
Implications:
The 20-gauge Arrow epidural catheters had similar tensile strengths as the 19-gauge epidural catheters but fractured at the traction site rather than at the fixed site. Thus, the 20-gauge Arrow catheter may be a reasonable alternative to the 19-gauge Arrow catheter.
More Related Videos
10:47Tri-layered Electrospinning to Mimic Native Arterial Architecture using Polycaprolactone, Elastin, and Collagen: A Preliminary Study
Published on: January 4, 2011
07:21Clinical Application of 24 G Cannula Needle and 3-0 Polypropylene Suture in Vas Deferens Exploration
Published on: February 10, 2023