Clinical experience with heat sterilization for reprocessing dialyzers.

A M Kaufman1, S Frinak, R O Godmere

  • 1Department of Medicine, Beth Israel Medical Center, New York, New York 10128.

ASAIO Journal (American Society for Artificial Internal Organs : 1992)
|July 1, 1992
PubMed
Summary

Heat sterilization is a safe and effective method for reprocessing dialyzers, offering advantages over chemical germicides. This method demonstrated no significant decrease in dialyzer performance or increase in adverse patient events during clinical trials.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Fermionic quantum processing with programmable neutral atom arrays.

Proceedings of the National Academy of Sciences of the United States of America·2023
Same author

Unraveling the relationship between mortality, hyponatremia, inflammation and malnutrition in hemodialysis patients: results from the international MONDO initiative.

European journal of clinical nutrition·2016
Same author

Entangling two transportable neutral atoms via local spin exchange.

Nature·2015
Same author

Association between pre hemodialysis serum sodium concentration and blood pressure: results from a retrospective analysis from the international monitoring dialysis outcomes (MONDO) initiative.

Journal of human hypertension·2015
Same author

Two-particle quantum interference in tunnel-coupled optical tweezers.

Science (New York, N.Y.)·2014
Same author

Effect of changes in the intravascular volume during hemodialysis on blood viscoelasticity.

Indian journal of nephrology·2011

Area of Science:

  • Nephrology
  • Biomedical Engineering
  • Sterilization Technologies

Background:

  • Chemical germicides are commonly used for dialysis reprocessing.
  • Heat sterilization presents potential advantages over chemical methods for dialyzer reprocessing.

Purpose of the Study:

  • To evaluate the safety and efficacy of heat sterilization for reprocessing polysulfone dialyzers.
  • To compare the performance and clinical outcomes of heat-sterilized dialyzers with those sterilized by chemical methods.

Main Methods:

  • Polysulfone dialyzers (Fresenius 60M or 80M) were heat sterilized at 105°C for 20 hours.
  • Clinical trials involved 180 patients undergoing 9,000 treatments with heat-sterilized dialyzers.
  • In vitro and in vivo performance metrics (clearances, KoA, Kt/V) were assessed at baseline and after multiple uses.

Related Experiment Videos

  • Dialyzer integrity and clinical blood leak incidence were monitored.
  • Main Results:

    • No pyrogenic reactions, sepsis, or subjective symptoms were reported.
    • In vitro clearances and in vivo KoA values showed no significant decrease after multiple uses.
    • Kt/V for urea remained comparable to baseline formaldehyde-sterilized dialyzers.
    • Dialyzer failure modes included integrity test failures (70%), blood leaks (8%), and reaching maximum usage limit (12%).
    • Clinical blood leaks occurred in less than 0.5% of treatments.

    Conclusions:

    • Heat sterilization is a safe and effective method for reprocessing polysulfone dialyzers.
    • Quality control is essential for the heat sterilization process.
    • Heat sterilization offers a promising alternative to chemical disinfection for dialysis reprocessing.