Related Experiment Video
Updated: Aug 30, 2026

Multi-Modal Home Sleep Monitoring in Older Adults
Published on: January 26, 2019
Assessment of a wrist-worn device in the detection of obstructive sleep apnea
Najib T Ayas1, Stephen Pittman, Mary MacDonald
1Department of Medicine, Brigham and Women's Hospital, Division of Sleep Medicine, Harvard Medical School, Boston, MA 02115, USA.
Objective:
To assess the accuracy of a wrist-worn device (Watch_PAT100) to diagnose obstructive sleep apnea (OSA).
Methods:
Thirty adult subjects with and without suspected OSA simultaneously had a standard in-laboratory polysomnogram (PSG) and wore the Watch_PAT100 during a full-night recording. PSG sleep and respiratory events were scored according to standard criteria. Watch_PAT data were analyzed with an automated computerized algorithm which calculated the frequency of respiratory events per hour of actigraphy measured sleep using a combination of peripheral arterial tonometry (PAT) signal attenuation, desaturation on pulse oximetry, and changes in heart rate. This yielded a PAT apnea hypopnea index (AHI).
Results:
Mean age was 47.0+/-14.8 years, mean body mass index 31.0+/-7.6 kg/m(2), mean PSG AHI 23+/-23.9 events per hour, and mean PAT AHI 23+/-15.9 events per hour. There was a significant correlation between PAT AHI and AHI by PSG (r=0.87, P<0.001). To assess sensitivity and specificity of Watch_PAT, we constructed receiver operator characteristic curves using a variety of AHI threshold values (10, 15, 20, and 30 events per hour). Optimal combinations of sensitivity and specificity for the various thresholds were 82.6/71.4, 93.3/73.3, 90.9/84.2, and 83.3/91.7, respectively.
Conclusions:
The Watch_PAT is a device that can detect OSA with reasonable accuracy. Thus, the Watch_PAT may be a useful method to diagnose OSA.
Related Concept Videos
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of apical pulse
Assessing the apical pulse is a critical nursing procedure, particularly indicated for:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...