Tympanometry versus direct middle ear pressure measurement in an artificial model: is tympanometry an accurate method
1Ear Research Laboratory, Department of Bio-Engineering, Tel-Aviv University, Israel. udicin@hotmail.co.il
Hypothesis:
Indirect tympanometric pressure measurements of the middle ear (ME) do not correspond to direct pressure measurements.
Background:
Tympanometry is an accepted method for estimating ME pressures. These pressure values are different, however, when measurements are made directly.
Materials And Methods:
Negative pressures were induced by volume changes in an artificial ME model with various "mastoid" volumes. Both tympanometric and direct pressure measurements were obtained simultaneously.
Results:
A substantial difference was demonstrated between tympanometric and direct pressure measurements. There was a linear correlation between tympanometry and direct pressures between -46.8 and -93.6 mmH2O, with a tympanometric overestimation of 40% to 20%, respectively. The smaller the "mastoid" volume, the greater the overestimation. The correlation was not linear for direct pressures between zero and -46.8 mmH2O. Tympanometric overestimation was greatest (i.e., 4-14-times the actual pressure) in the pressure range between zero and -12 mmH2O.
Conclusions:
Tympanometry does not yield precise ME pressure values. In a model with a larger "mastoid" and an actual pressure range of -46.8 to -93.6 mmH2O tympanometric readings are somewhat less unreliable (error of 40%-20%). However, tympanometry is in error of more than 400% in a model with a small mastoid in which the actual ME pressure is only slightly negative (i.e., range between -3.9 and -46.8 mmH2O). Therefore, this study implies that tympanometry has little use for measuring ME pressure especially in chronic ears or their sequelae that are usually characterized by having small negative pressures and small mastoids.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sound as Pressure Waves
The pressure fluctuation depends on the difference in displacements between the successive points in the...
Definition and Measurement of Pressure: Atmospheric Pressure, Barometer, and Manometer
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...