Using the frame averaging of aS500 EPID for IMRT verification
1Medical Physics Department, Memorial Sloan-Kettering Cancer Center, 1275 York Avenue, New York, New York 10021, USA. changj@mskcc.org
Journal of Applied Clinical Medical Physics
|November 8, 2003
Summary
This study validates the aS500 Electronic Portal Imaging Device (EPID) for intensity-modulated radiation therapy (IMRT) beam delivery verification. Simple correction methods ensure accurate dose profile and central axis dose measurements for clinical use.
Area of Science:
- Medical Physics
- Radiation Oncology
- Imaging Technology
Background:
- Intensity-modulated radiation therapy (IMRT) requires precise beam delivery verification.
- Electronic Portal Imaging Devices (EPIDs) are used for quality assurance in radiation therapy.
- The aS500 EPID's synchronous, frame-averaging acquisition mode presents potential challenges for accurate dose measurement.
Purpose of the Study:
- To evaluate the efficacy of the aS500 EPID for IMRT beam delivery verification.
- To assess and correct for potential errors in the synchronous, frame-averaging acquisition mode.
- To determine the accuracy of the aS500 EPID for measuring dose profiles and central axis dose.
Main Methods:
- Utilized synchronous, frame-averaging acquisition with the aS500 EPID.
- Developed and applied correction schemes for EPID dead time and data gaps.
- Converted averaged EPID images to dose profiles using linear calibration.
- Compared measured dose profiles and central axis dose with planned values using linear regression analysis for 25 prostate IMRT fields.
Main Results:
- Identified and corrected for insignificant data gaps and significant EPID dead time for clinical Monitor Units (MU >100).
- Achieved excellent agreement between measured and planned profiles (mean sigma 1.9%, SD 0.5%) after applying correction methods.
- Demonstrated agreement in central axis dose better than 2.0%.
Conclusions:
- The aS500 EPID, with implemented correction schemes, is effective for verifying IMRT beam delivery for clinical MU (>100).
- Satisfactory results can be achieved with the current configuration, even before vendor-specific upgrades.
- This validates the use of the aS500 EPID as a reliable tool for IMRT quality assurance.