Prediction of the apnea-hypopnea index from overnight pulse oximetry
Ulysses J Magalang1, Jacek Dmochowski, Sateesh Veeramachaneni
1Division of Pulmonary, Critical Care, and Sleep Medicine, University at Buffalo, New York, USA.
Study Objective:
s: To compare the relative usefulness of the different indexes derived from pulse oximetry in the diagnosis of obstructive sleep apnea (OSA), and to determine if a combination of these indexes improves the prediction of the apnea-hypopnea index (AHI) measured by polysomnography.
Design:
Prediction model developed from 224 patients, validated prospectively in 101 patients from the same center (group 1) and in 191 patients from a different sleep center (group 2).
Setting:
Two independent sleep clinics run by university sleep specialists.
Participants:
Patients who underwent polysomnography for suspicion of OSA.
Interventions:
The following indexes were calculated from pulse oximetry recordings performed simultaneously during polysomnography: (1) Delta index, the average of the absolute differences of oxygen saturation between successive 12-s intervals; (2) desaturation events per hour to 2%, 3%, and 4% levels; and (3) cumulative time spent below 90%, 88%, 86%, 84%, 82%, and 80% saturation.
Measurements And Results:
The best predictor was the Delta index, although desaturation events provided similar levels of diagnostic accuracy. An aggregation of multivariate models using combination of indexes reduced the prediction error (r(2) = 0.70) significantly (p < 0.05) compared to using the Delta index alone (r(2) = 0.60). The proportion of subjects from the validation groups within 95% confidence interval (CI) of the derivation group was 90% (95% CI, 83 to 95%) and 91% (95% CI, 86 to 95%) for groups 1 and 2, respectively. The overall likelihood ratios for the aggregated model in all patient groups were 4.2 (95% CI, 3.3 to 15.3), 3.4 (95% CI, 2.7 to 4.3), 3.0 (95% CI, 2.2 to 4.1), and 6.7 (95% CI, 4.9 to 9.2) for normal (AHI < 5/h), mild (AHI 5 to < 15/h), moderate (AHI 15 to < 30/h), and severe (AHI > or = 30/h) disease, respectively.
Conclusions:
The Delta index and oxygen desaturation indexes provided similar levels of diagnostic accuracy. The combination of indexes improved the precision of the predicted AHI and may offer a potentially simpler alternative to polysomnography.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Respiratory Assessment: Purpose and Indications
Objectives and Importance:
The primary goal of respiratory assessment is to evaluate patients at early risk of clinical deterioration. Since respiratory distress often precedes other signs of declining health, breathing patterns and sounds become a...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Respiratory Volumes and Capacities I