Related Experiment Video
Updated: Aug 9, 2026

Imaging Odor-Evoked Activities in the Mouse Olfactory Bulb using Optical Reflectance and Autofluorescence Signals
Published on: October 31, 2011
Optical bio-sniffer for ethanol vapor using an oxygen-sensitive optical fiber
Kohji Mitsubayashi1, Takuo Kon, Yuki Hashimoto
1Department of Human and Information Science, School of Information Technology and Electronics, Tokai University, 1117 Kitakaname, Hiratsuka, Kanagawa 259-1292, Japan. mitsu@keyaki.cc.u-tokai.ac.jp
Abstract:
An optical bio-sniffer for ethanol was constructed by immobilizing alcohol oxidase (AOD) onto a tip of a fiber optic oxygen sensor with a tube-ring, using an oxygen sensitive ruthenium organic complex (excitation, 470 nm; fluorescent, 600 nm). A reaction unit for circulating buffer solution was applied to the tip of the device. After the experiment in the liquid phase, the sniffer-device was applied for gas analysis using a gas flow measurement system with a gas generator. The optical device was applied to detect the oxygen consumption induced by AOD enzymatic reaction with alcohol application. The sensor in the liquid phase was used to measure ethanol solution from 0.50 to 9.09 mmol/l. Then, the bio-sniffer was calibrated against ethanol vapor from 0.71 to 51.49 ppm with good gas-selectivity based on the AOD substrate specificity. The bio-sniffer with the reaction unit was also used to monitor the concentration change of gaseous ethanol by rinsing and cleaning the fiber tip and the enzyme membrane with buffer solution.