Reproducibility of brain ADC histograms
Stefan C A Steens1, Faiza Admiraal-Behloul, Jorrit A Schaap
1Department of Radiology, C2S, Leiden University Medical Center, P.O. Box 9600, 2300 RC, Leiden, The Netherlands. s.c.a.steens@lumc.nl
European Radiology
|November 18, 2003
Summary
Apparent diffusion coefficient (ADC) histogram parameters vary with diffusion-weighted imaging (DWI) acquisition techniques. However, ADC histogram analysis is reproducible, and FLAIR DWI reduces confounding effects from cerebrospinal fluid.
Area of Science:
- Radiology
- Neuroimaging
- Medical Physics
Background:
- Apparent diffusion coefficient (ADC) histogram analysis is a quantitative method for evaluating diffusion-weighted imaging (DWI) data.
- Understanding the impact of acquisition parameters on ADC histograms is crucial for consistent interpretation.
- Scan-rescan reproducibility is essential for the clinical utility of any imaging biomarker.
Purpose of the Study:
- To evaluate how different DWI acquisition techniques affect whole-brain ADC histogram parameters.
- To assess the scan-rescan reproducibility of ADC histogram analysis.
- To investigate the utility of FLAIR DWI in mitigating confounding factors.
Main Methods:
- Healthy subjects underwent DWI with varying maximum b-values (0-800, 0-1000, 0-1500 s/mm²) and FLAIR DWI (0-1000 s/mm²).
- Sequences were repeated with and without repositioning to assess reproducibility.
- Paired-sample t tests were used to compare histogram parameters (peak location, peak height, mean ADC) and region of interest (ROI) mean ADC.
Main Results:
- Increasing maximum b-values shifted ADC histograms to lower values with increased peak height (p<0.01).
- FLAIR DWI resulted in lower ADC values and a significantly higher, narrower peak (p<0.01) compared to standard DWI.
- No significant differences were found in ROI mean ADC or scan-rescan reproducibility across techniques.
Conclusions:
- DWI acquisition techniques significantly influence ADC histogram parameters.
- ADC histogram analysis demonstrates robustness and reproducibility with a given pulse sequence.
- FLAIR DWI effectively reduces the partial-voluming effect of cerebrospinal fluid, improving histogram analysis reliability.