Measurement of ultrasonic attenuation coefficient in polymer gel dosimeters
Melissa L Mather1, Paul H Charles, Clive Baldock
1School of Physical and Chemical Sciences, Queensland University of Technology, Brisbane, Australia.
Abstract:
A technique is described for investigation of the ultrasonic attenuation coefficient for evaluation of absorbed dose in polymer gel dosimeters. Using this technique the attenuation coefficient as a function of absorbed dose in PAG and MAGIC polymer gel dosimeters was measured. The ultrasonic attenuation coefficient dose sensitivity for PAG was found to be 2.9 +/- 0.3 dB m(-1) Gy(-1) and for MAGIC gel 4.2 +/- 0.3 dB m(-1) Gy(-1). Unlike previous studies of ultrasonic attenuation in polymer gel dosimeters this technique enables a direct measure of the attenuation coefficient.