Related Experiment Video
Updated: Aug 30, 2026

20 mJ, 1 ps Yb:YAG Thin-disk Regenerative Amplifier
Published on: July 12, 2017
Output energy changes of quartz contact probe for Er:YAG laser with tooth ablation
Koichiro Yamazaki1, Toru Eguro, Toru Maeda
1Department of Endodontics and Operative Dentistry, School of Dentistry at Tokyo, Nippon Dental University, 2-3-16 Fujimi, Chiyoda-ku Tokyo, 102-8158 Japan.
Abstract:
The purpose of this study was to examine the output energy changes of the contact probe for Er:YAG laser with tooth ablation and evaluate the effect of contact surface polishing on output energy. The Er:YAG laser was irradiated to enamel of extracted human incisors for 100 min (5 min x 20 times). The output energy of contact probes was measured every 5 min (n = 3). After 100 min, the contact surfaces were polished using silicon carbide paper up to #1500, and the output energy of the probe was measured. The contact surfaces of the probes were observed and analyzed using an EPMA. After 100 min, the output energy showed 49.4% of the energy of an unused probe. After the contact surfaces were polished to #1500, the output energy returned to 96.4% of baseline. It is suggested that the output energy of the contact probe decreased with tooth ablation, and polishing of the contact surface was effective to regain the energy.