Reusability of plastic scintillators used in beta emitter activity determination
A Tarancón1, J F García, G Rauret
1Departament de Química Analítica, Universitat de Barcelona, Diagonal 647, E-08028 Barcelona, Spain. alex.tarancon@apolo.qui.ub.es
Abstract:
The use of plastic scintillation in the measurement of beta emitter activity is a suitable alternative to classical detection techniques in spite of the high cost and potential memory effects. Reuse of the plastic scintillator beads after a cleanup procedure would overcome both drawbacks. In the present paper, we tested several cleaning solutions (water, HCl 0.1M, commercial soap, EDTA 0.01 M pH=3, citric acid 0.1 M pH=3). The recommended procedure consists of the immersion of the beads in three consecutive citric acid baths for 24 h. If this protocol is followed, plastic beads contaminated with 90Sr/90Y, 14C, and 3H recover their initial condition, and can be used in successive determinations.