Related Experiment Video
Updated: Aug 30, 2026

NiO Nanoflowers for Non-Enzymatic Amperometric Detection of Glucose
Published on: December 30, 2025
A nitric oxide biosensor based on the multi-assembly of hemoglobin/montmorillonite/polyvinyl alcohol at a pyrolytic
Jiantao Pang1, Chunhai Fan, Xinjian Liu
1Department of Biochemistry and National Key Laboratory of Pharmaceutical Biotechnology, Nanjing University, Nanjing 210093, China.
Abstract:
Direct electron transfer of hemoglobin (Hb) can be achieved in a Hb/montmorillonite (MMT)/polyvinyl alcohol multi-assembly at a pyrolytic graphite (PG) electrode. Accordingly, a novel nitric oxide (NO) biosensor is proposed. The reduction of NO is observed at a potential of approximately -783 mV (vs. SCE) at pH 5.5. At optimum pH, this biosensor shows a wide linear range of 1.0x10(-6)-2.5x10(-4) mol/l with a detection limit of 5.0x10(-7) mol/l. The sensor-to-sensor reproducibility is good consideringmacr; a relative standard deviation of 3.5% in five independent determinations at 5.0x10(-5) mol/l NO. The modified electrode is conveniently constructed and durable in long-term operations.