Accuracy and applicability of the Terumo ES-H55 double-cuff sphygmomanometer for hospital use
Osamu Tochikubo1, Kiyoko Nishijima, Kenji Ohshige
1Department of Public Health, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama 236-0004, Japan. tocchi@med.yokohama-cu.ac.jp
Background:
We have developed a new blood pressure (BP)-measuring device with a double-cuff, and evaluated its accuracy and applicability for practical use.
Design:
The double-cuff method features a small cuff placed in the centre of the compression cuff to detect the oscillation of arterial pulsation, by which algorithms for the determination of systolic BP (SBP) and diastolic BP (DBP) were rendered more objective than with the conventional oscillometric method. Algorithms for automatic BP determination were developed from the oscillation data taken with the small cuff from 217 men and 256 women, and then installed in the Terumo ES-H55 device (length: 17 cm, weight: 120 g). This ES-H55 device was tested on 87 subjects (65 hypertensives and 22 normotensives) to compare its accuracy with that of the auscultatory method according to the AAMI SP-10 protocol.
Results:
Using Bland and Altman scatter plots, the difference in SBP between the ES-H55 device and the auscultatory method was within +/- 5 mmHg in 88% of subjects, and within +/- 10 mmHg in 97% (mean difference +/- SD: 0.2 +/- 3.9 mmHg). The difference in DBP was within +/- 5 mmHg in 80%, and within +/- 10 mmHg in 97% (difference: -0.1 +/- 4.6 mmHg). These results satisfied AAMI guidelines, and this device was easy to use.
Conclusions:
The ES-H55 device is compact and light, with clear algorithms for the determination of BP, and satisfies AAMI criteria, so that it is considered a useful BP-measuring device that could be used instead of a mercury sphygmomanometer by nurses and doctors in hospital wards.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...