Effect of oral contrast agents on computed tomography-based positron emission tomography attenuation correction in
Gerald Antoch1, Walter Jentzen, Lutz S Freudenberg
1Department of Diagnostic and Interventional Radiology, University Hospital Essen, Germany. gerald.antoch@uni-essen.de
Rationale And Objectives:
To evaluate the effect of iodine- and barium-based contrast agents on the computed tomography (CT)-based positron emission tomography (PET) attenuation correction in dual-modality PET/CT.
Methods:
Experiments were conducted on a Society of Nuclear Medicine/National Electrical Manufacturers Association-PET phantom equipped with cylinders containing [18F]-2-fluoro-2-desoxy-D-glucose. The main compartment was filled with iodine (0.5-10%), barium (0.5-50%), or water (negative control). The error in attenuation correction was determined by comparison of measured tracer quantities in the presence of contrast agents with expected quantities. Contrast agent attenuation was demonstrated to be comparable to in vivo conditions.
Results:
The presence of contrast agents resulted in an overestimation of the intracylindrical activity concentration on PET images and overestimation directly related to contrast concentrations (iodine 5-38%; barium 15-580%). Iodine and barium concentrations in clinical use resulted in an activity overestimation of 20 +/- 1.8% for iodine and 21 +/- 2.9% for barium.
Conclusion:
An overestimation of the tracer activity concentration is to be expected in the presence of oral contrast agents, if PET attenuation correction is attained CT-based.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET