Gastroscope processing in washer-disinfectors at three different temperatures.

B Zühlsdorf1, A Winkler, B Dietze

  • 1Freie Universität Berlin, Universitätsklinikum Benjamin Franklin, Technische Hygiene, Hindenburgdamm 27, 12203 Berlin, Germany.

The Journal of Hospital Infection
|November 25, 2003
PubMed
Summary

Processing gastroscopes at higher temperatures significantly reduces bacterial contamination. Increasing endoscope washer-disinfector temperature from 43°C to 56°C ensures patient safety and adequate reprocessing.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Experimental Distributed Quantum Sensing in a Noisy Environment.

Physical review letters·2025
Same author

Photon-Interfaced Ten-Qubit Register of Trapped Ions.

Physical review letters·2025
Same author

Rapid cortical auditory evoked potentials audiometry.

International journal of audiology·2025
Same author

Influence of the Scalpel Finishing Technique on Marginal Gap Formation in Class II Resin Composite Restorations.

Operative dentistry·2022
Same author

Efficacy of five 'sporicidal' surface disinfectants against Clostridioides difficile spores in suspension tests and 4-field tests.

The Journal of hospital infection·2022
Same author

Interceptive use of pendulum for advance resolution of class II patient: a guided eruption plan to reduce stress on permanent teeth root.

Journal of biological regulators and homeostatic agents·2021

Area of Science:

  • Microbiology
  • Infection Control
  • Medical Device Reprocessing

Background:

  • Endoscopes require thorough reprocessing to prevent patient-to-patient transmission of infections.
  • Current chemo-thermal reprocessing in Germany uses approximately 56°C.
  • Variations in reprocessing temperatures may impact the efficacy of bacterial load reduction.

Purpose of the Study:

  • To evaluate the microbiological effectiveness of endoscope washer-disinfectors at different processing temperatures.
  • To determine the optimal temperature for reprocessing gastroscopes to minimize residual contamination.
  • To assess the impact of temperature on bacterial reduction and endoscope material integrity.

Main Methods:

  • Hygienic and microbiological testing of 87 gastroscopes.

Related Experiment Videos

  • Processing at washer-disinfector temperatures of 43°C, 51°C, and 56°C.
  • Microbiological analysis of suction/biopsy channels using culture media and membrane filtration.
  • Main Results:

    • Bacterial counts exceeded German guideline levels at all tested temperatures.
    • A significant reduction in aerobic bacteria, Gram-negative bacilli, and pseudomonads was observed when increasing temperature from 43°C to 51°C.
    • Further significant reductions in aerobic bacteria and pseudomonads were noted with a temperature increase from 51°C to 56°C.

    Conclusions:

    • Lowering reprocessing temperatures risks patient safety due to inadequate bacterial load reduction.
    • Endoscope washer-disinfectors must meet current draft standards for effective reprocessing.
    • Maintaining higher processing temperatures is crucial for ensuring hygienic endoscope processing and patient safety.