Monolayer-protected gold nanoparticles as a stationary phase for open tubular gas chromatography
Gwen M Gross1, David A Nelson, Jay W Grate
1Center for Process Analytical Chemistry, Department of Chemistry, Box 351700, University of Washington, Seattle, Washington 98195, USA.
Abstract:
The use of a thin film of monolayer-protected gold nanoparticles (MPNs) as a stationary phase for gas chromatography (GC) is reported. Deposition of a MPN film was obtained in a 2-m, 530-microm-i.d. deactivated silica capillary using gravity to force the solution containing the MPN material through the capillary. By SEM analysis, the average film thickness was determined to be 60.7 nm. The retention behavior for the dodecanethiol MPN column was studied using four compound classes (alkanes, alcohols, aromatics, ketones), and retention orders were objectively compared to a commercially available column (AT-1, 100-nm film thickness). Separation of an eight-component mixture was performed using both isothermal and temperature-programming methods with the dodecanethiol MPN phase and compared to an isothermal separation with the AT-1 phase. The AT-1 phase separation had an efficiency, N, of 6200 (k' = 0.33) while the dodecanethiol MPN phase separation had an efficiency, N, of 5700 (k' = 0.21) for the same analyte, octane. The reduced plate height, h, for octane was found to be less than 1 at the optimum linear flow velocity, indicating the MPN column operated near the optimum possible performance level. Robustness of the MPN phase is also discussed with consistent performance observed over several months. Overall, MPNs appear promising as a stationary-phase material for GC and as an experimental platform to study their thermodynamic and mass-transfer properties.
Related Concept Videos
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...