Related Experiment Video
Updated: Jan 13, 2026

Precise Electrochemical Sizing of Individual Electro-Inactive Particles
Published on: August 4, 2023
Zeptomole voltammetric detection and electron-transfer rate measurements using platinum electrodes of nanometer
John J Watkins1, Jinyuan Chen, Henry S White
1Department of Chemistry, University of Utah, 315 South 1400 East, Salt Lake City, Utah 84112, USA.
Abstract:
The characterization of quasi-hemispherical Pt electrodes of nanometer dimensions (radius 2-150 nm), prepared by electrophoretic coating of etched Pt wires with poly-(acrylic acid), is described. The goals of these experiments are to estimate the accuracy of using steady-state voltammetric limiting currents (i(lim)) in determining the true electrode area and to develop new electrochemical methods for rapidly screening individual electrodes for non-ideal geometries. Electrochemical active areas were determined by measuring the electrical charge (Q) associated with oxidation of adsorbed bis(2,2'-bipyridine)chloro(4,4'-trimethylenedipyridine)osmium(II) in fast-scan voltammetric measurements (scan rate 1000 V/s). Voltammetric peaks corresponding to oxidation of as few as approximately 7000 molecules (approximately 11 zmol) at individual electrodes are reported, allowing precise measurement of electrode areas as small as approximately 10(-10) cm2. A plot of i(lim) (for a soluble redox species) versus Q1/2 (for an adsorbed redox species), constructed from i(lim)-Q1/2 data pairs obtained as a function of the electrode radius, is shown to be linear if the electrode geometry is independent of electrode radius; departure of experimental values from the straight-line plot is a diagnostic indicator of a nonideal electrode geometry. The results indicate that approximately 50% of the electrodes prepared by the electrophoretic polymer-coating procedure are quasi-hemispherical, the remaining being recessed slightly below the polymer coating. The heterogeneous electron-transfer rate constant for the oxidation of the ferrocenylmethyltrimethylammonium cation in H2O/ 0.2 M KCl was also determined from steady-state voltammetry using the method of Mirkin and Bard and found to be 4.(8) +/- 3.(2) cm/s with alpha = 0.6(4) +/- 0.1(5).
Related Concept Videos
Voltammetry: Factors Affecting Measurements
Voltammetric Techniques: Pulse Voltammetry
Voltammetry: Overview
A voltammetric cell uses three electrodes: a working electrode, a reference electrode, and an auxiliary electrode. The redox reactions occur in the working...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...
Voltammetric Techniques: Linear-Scan (E vs Time)
Potentiometry: Membrane Electrodes