[Navigated reposition of transverse acetabulum fractures. A precision analysis]
Related Experiment Video
Updated: Aug 30, 2026

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
1Unfallchirurgische Klinik, Medizinische Hochschule, Hannover. huefner.tobias@mh-hannover.de
Abstract:
Up to now navigated reduction control based on computed tomography (CT) image data could not be used commercially. With newly developed software, a transverse fracture of the acetabulum was reduced with navigation control in a laboratory test. The results were compared to visual and tactile control in a foam pelvis and specimen. Measurements were done with another magnet-based navigation system. The residual dislocation was measured with translation (mm) and rotation (degrees). Compared with visually controlled reduction, navigated reduction led to a residual dislocation of 0.7 mm and 0.9 degrees. Navigated reduction based on CT image data is also accurate for reduction of joint fractures under laboratory conditions. Further improvements of the software are planned for later in vivo use.