Evaluation of modulators and electron-capture detectors for comprehensive two-dimensional GC of halogenated organic
E Maria Kristenson1, Peter Korytár, Conny Danielsson
1Department of Analytical Chemistry and Applied Spectroscopy, Faculty of Sciences, Vrije Universiteit, de Boelelaan 1083, 1081 HV Amsterdam, The Netherlands. kristens@few.vu.nl
Abstract:
Different cryogenic and a heated GC x GC modulator(s) were evaluated and compared for the analysis of high-boiling halogenated compounds. The cryogenic modulators investigated were: (i) the longitudinally modulated cryogenic system; (ii) the liquid-nitrogen-cooled jet modulator (KT2001); (iii) a dual-jet CO2 modulator (made in-house); (iv) a semi-rotating cryogenic modulator (made in-house) and (v) a CO2 loop modulator (KT2003); the heated modulator was the slotted heater system (sweeper). Each modulator was optimised with respect to analyte peak widths at half height in the second-dimension. n-Alkanes, chlorinated alkanes, polychlorinated biphenyls (PCBs) and fluorinated polycyclic aromatic hydrocarbons (F-PAHs) were used as test analytes. The flow rate of the coolant was found to be an important parameter, i.e. the flow rate of the gaseous nitrogen in the KT2001, and of the liquid CO2 in the other cryogenic modulators. For the slotted heater the stroke velocity and pause time were important parameters. This modulator had a limited application range in terms of temperature due to a necessary 100 degrees C difference between sweeper and oven temperature. All cryogenic modulators were found to be suitable for the GC x GC analysis of high-boiling compounds, but the CO2 modulators are to be preferred to the KT2001 due to a wider application range and slightly narrower peaks. As regards the performance of three commercially available electron-capture detectors (ECDs), the aim was to obtain narrow peak widths in GC x GC, i.e. to avoid band broadening caused by the cell volume. The most important parameters were the flow rate of the make-up gas and the detector temperature which both should be as high as possible. Comparison of analyte peak widths obtained with ECD mode and flame ionisation detection (FID) showed that all ECDs exhibited band broadening compared to the FID. The narrowest peaks were obtained with the Agilent micro-ECD, which has a cell volume of only 150 microl.
More Related Videos
10:14Chromatographic Fingerprinting by Template Matching for Data Collected by Comprehensive Two-Dimensional Gas Chromatography
Published on: September 2, 2020
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.