Determination of Cardiac Output in a Porcine Model for Ex Vivo Pulmonary Perfusion
Published on: June 28, 2024
A comparative evaluation of thermodilution and partial CO2 rebreathing techniques for cardiac output assessment in
Monica Rocco1, Gustavo Spadetta, Andrea Morelli
1Department of Anesthesiology and Intensive Care, University of Rome La Sapienza, Viale del Policlinico 155, 00161 Rome, Italy. monica.rocco@uniroma1.it
Objective:
To evaluate the reliability and clinical value of partial noninvasive CO2 (NICO2) rebreathing technique for measuring cardiac output compared with standard thermodilution in a group of intensive care nonpostoperative patients.
Design And Setting:
Clinical investigation in a university hospital ICU.
Patients:
Twelve mechanically ventilated patients with high (n=6) and low (n=6) pulmonary shunt fractions.
Measurements And Results:
Thirty-six paired measurements of cardiac output were carried out with NICO2 and thermodilution in patients ventilated in pressure-support mode and sedated with a sufentanil continuous infusion to obtain a Ramsay score value of 2. The mean cardiac output was: thermodilution 7.27+/-2.42 l/min; NICO2 6.10+/-1.66 l/min; r2 was 0.62 and bias -1.2 l/min+/-1.5. Mean values of cardiac output were similar in the low shunt group (Qs/Qt < 20), with r2=0.90 and a bias of 0.01 l/min+/-0.4; conversely, in the high pulmonary shunt group (Qs/Q > 35%) the mean was 9.32+/-1.23 l/min with thermodilution and a mean NICO2CO value was 6.97+/-1.53 l/min, with r2 of 0.38 and a bias of -2.3 l+/-1.2 min.
Conclusions:
The partial CO2 rebreathing technique is reliable in measuring cardiac output in nonpostoperative critically ill patients affected by diseases causing low levels of pulmonary shunt, but underestimates it in patients with shunt higher than 35%.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: